APC AR1100 Manual do Operador Página 130

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 144
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 129
116
SPECIFICATIONS AND ACCESSORIES: Safety Features
5.2 Safety Features
To ensure safe use, the device is provided with the following safety features.
<Main body position detection switch>
The switch detects if the main body is fully pulled toward the operator.
Unless the switch is turned on, the switchover between the R/K measuring unit and NT mea-
suring unit is disabled.
<Patient sensor>
This sensor detects if the patient's chin is placed on the chinrest during the up-and-down
movement of the chinrest.
While the patient is being detected by the sensor, the chinrest is not initialized.
<Safety stopper>
For safety, this stopper provides a safety space so that the air nozzle does not touch a
patient’s eye during measurements.
The amount of space needed for safety depends on the patient. Change the position of the
stopper for each patient to ensure the proper amount of space for safety.
Vista de página 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 143 144

Comentários a estes Manuais

Sem comentários