OPERATING PROCEDURES: Selecting the Mode
2
2.4.1 Switching to manual mode
By pressing the manual mode button during measurements, it is possible to turn off both the
auto-tracking and auto-shooting functions (manual mode).
Press the manual mode button to turn off the auto-tracking function according to the state the
patient’s eye after starting the measurement with the auto-tracking function on. (The manual mode
button functions even when the auto-shooting function only is turned on.) Allowing the operator to
turn off the auto-tracking and auto-shooting functions with a touch of the button, the manual mode
button saves the operator the trouble of pressing the auto button repeatedly.
1 Press the manual mode button during
measurements.
The auto-tracking ON mark and auto-shooting
ON mark will be blanked ( ), indicating
that these functions are turned off.
2 Manually align the main body to the eye and bring the eye into focus.
3 Press the start button to start the measurement.
One of the following operations returns to the state before the manual mode button is pressed.
• Pressing the manual mode button again
• Pressing the CLR button
• Pressing the print button
• Switch the measurement part (RK measurement and NT measurement).
The functions assigned to the Auto button depend on the 28. TRACKING SW
parameter settings.
For the setting method, see “2.8 Parameter Settings” (page 76).
When the manual mode button is pressed to turn off the auto-tracking and auto-
shooting modes (manual mode), the auto-tracking ON mark and auto-shooting ON mark
will be blanked ( ).
For the eyes with a small pupil and with the corneal vertex shifted from the pupil center, if
the auto-tracking function is turned on and the pupil overlaps the min. pupil mark, the eye
may not be measured.
To measure such eyes, turn off the auto-tracking function and align the min. pupil mark with the pupil.