OPERATING PROCEDURES: Parameter Settings
7 : AR THUMBNAIL
Selects whether or not to display the thumbnail screen of the measurement ring image during AR mea-
surement.
When YES is selected, a thumbnail of the measurement ring image is displayed to the left of the
screen after AR measurement is complete. Press the ring image enlargement button for full
screen display of the ring image.
[KM (KM measurement)]
* Underlined options indicate factory settings.
KM UNIT (unit for corneal curvature radius)
Selects whether or not to express the corneal curvature radius obtained in KM measurement in radius
(mm) or in diopter (D).
8 : KM DISPLAY (representation of corneal curvature radius)
Selects the KM measurement representation between R1 (flattest meridian) and R2 (steepest merid-
ian), and AVE (averages of R1 and R2) and CYL (amount of corneal astigmatism).
9 : REF. INDEX (corneal refractive index)
Selects the corneal refractive index.
10 : KM CONTINUE (continuous KM measurement)
Selects the number of multiple KM measurements in KM measurement.
[NT (NT measurement)]
* Underlined options indicate factory settings.
21 : SET LOW CONF (Display of low confidence data)
Selection of whether or not to display low confidence data.
No. Parameter Settings
11 KM UNIT mm
/ D
12 KM DISPLAY R1, R2 / AVE, CYL
13 REF. INDEX 1.3320 / 1.3360 / 1.3375
14 KM CONTINUE 3 - 10 (Factory setting: 3)
No. Parameter Settings
21 SET LOW CONF YES / NO
22 LOW CONF LV. YES / NO
23 LOW CONF ALARM YES / NO
24 FIX LED BLINK YES / NO
25 NT CONTINUE 3 - 10 (Factory setting: 3)
26 DECIMAL DIGIT YES / NO
27 MEAS INTERVAL SHORT / NORMAL / LONG